Main Features 1. Large format display with the ability to switch between Chinese and English, simple and clear switching between screens, modern and aesthetic design. 2. The device is equipped with a 7-inch color LCD display. 3. Perfect intelligent multi-dimensional...
1. Large format display with the ability to switch between Chinese and English, easy and clear switching between screens, modern and aesthetic design.
2. The device is equipped with a 7-inch color LCD display.
3. Perfect intelligent multidimensional chromatography system based on a 32-bit ARM9 microcontroller and original foreign software; Wide range of heating rates, fast cooling system, highly stable temperature control technology.
4. Reliable protection against overheating, as well as an alarm about a break or short circuit of the platinum thermal resistance, preventing uncontrolled heating.
5. In the range of up to 180 °C, the accuracy of maintaining the temperature in the column thermostat reaches ±0.01 °C.
6. A developed self-diagnosis system that allows you to quickly determine the location and type of malfunction.
7. Optional: built-in analog-to-digital converter for direct digital signal output.
8. Ability to work with two TSDs, two FIDs, or with TSD and FID simultaneously.
9. Possibility of simultaneous installation of three detectors and a methane converter.
10. Optional: manual sample input, automatic start of the sampling device and other options.
11. Possibility of connecting external devices: sampler for headspace analysis, thermal desorber, automatic sampler for liquids, automatic gas dispenser, etc.
12. Function for saving parameters when the power is turned off (non-volatile memory).
Column thermostat temperature control range: temperature control is supported from –80 °C to +420 °C (under certain conditions).
Temperature maintenance accuracy: no worse than ±0.1 °C.
Temperature programming: 10 adjustable levels.
Temperature gradient: ±1 °C (when programming the temperature in the range of 100–360 °C).
Heating rate: from 0.1 to 40 °C/min (maximum speed – 40 °C/min).
Temperature control range of the evaporator and detectors: from room temperature +10 °C to 400 °C.
Supply voltage: 220 V ±10%; maximum power consumption: 2200 W.
Flame Ionization Detector (FID)
Detection limit: Mt≤6×10-12g/s
Noise: ≤4×10-14A
Drift: ≤2×10-13A/30min
Linear range: ≥ 10⁷
Dimensions and weight
External dimensions: 560 × 460 × 450 mm (length × width × height)
Device weight: 40 Kg