Operating modes 1. Degassing for chromatography: temperature 50 °C; shaking 20 min; Sludge 10 min. 2. Water-soluble acids: temperature 75 ° C; shaking 5 min; settling time 0 min. 3. Other physical shaking: temperature from 0 to 110 °C (free setting); shaking from 0 to...
1. Degassing for chromatography: temperature 50 °C; shaking 20 min; Sludge 10 min.
2. Water-soluble acids: temperature 75 ° C; shaking 5 min; settling time 0 min.
3. Other physical shaking: temperature from 0 to 110 °C (free setting); shaking from 0 to 99 min (free setting); settling time from 0 to 99 min (free setting).
4. Manual shaking start: Start or stop shaking manually.
1. Device screen: 5-inch color touch screen with full animation of the entire heating and oscillation process. Switch between Chinese and English interfaces at the touch of a button.
2. Temperature control range: room temperature - 100 °C
3. Temperature control accuracy: room temperature - 50°C ±0.2°C,50°C - 100°C ±0.3°C
4. Temperature control method: digital automatic PID control
5. Oscillation amplitude: 35mm
6. Oscillation frequency: 275±3 times/min
7. Method of placing samples: 8 syringes with a volume of 100 ml (two tiers)
8. 4 triangular flasks with a volume of 250 ml (optional);
9. Maximum power consumption: 500VA
10. Operating voltage: AC220V±20%
11. Power frequency: 50Hz±5%
12. Ambient temperature: 10°C to 40°C;
13. Ambient humidity: ≤85%
14. Weight: 35Kg